  Fully Automated Haematology Analyser
 
 
X-ray & Dignistic Equipments
 
 
 

 

   

Click To Enlarge

Fully Automated Haematology Analyser

3 Part differential Cell Counters

General Technical Specifications Principles Electrical resistance for counting and SFT method for hemoglobin. Parameters WBC, Lymph, Mid, Gran, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, WBC Histogram, RBC Histogram, PLT Histogram, P-LCR. Working mode Single counting pathway + Independent HGB detection system Operating mode Color Windows interface, figure button, mouse input Specification 23 parameters, with 3 part WBC differntial Aperture Diameter 70um 10 uL sampler volume Memory for 30,000 results including histograms Throughput: 35 samples/hour Automatic and manual calibration Automatic dilution when power on/off High voltage cleaning after each measurement QC mode: X-B, L-J curve, SD, CV Interface 1 parallel interface, 2 USB

General Technical Specifications

Principles

Electrical resistance for counting and SPT methods for hemoglobin.

Parameters

WBC, Lymph, Mid Gran, Lymph %, Mid %, Gran %, RBC, HGB. HCT, MCV, MCH, MCHC, RDW-SD, PLT, MPV, PDW, PCT, WBC Histrogram, RBC Histrogram, PLT Histrogram, P-LCR.

Working mode

Double counting pathway + independent HGB detection systems

Operating mode

Color Windows interface, figure button, mouse input.

Interface

1 parallel interface, 2 USB interfaces, 2 RS232 interfaces (with net), 1 VGA interface.

Specification

> 23 parameters with 3 part WBC differential
> Aperture Diameter volume
> 10 ul sampler volume
> Memory for 30,000 results including histograms
> Throughput : 60 samples / hour
> Automatic and manual calibration
> Auomatic dilution when power on / off
> High voltage clednind after each measurement
> QC mode : X-B, L-J curve, SD, CV.

Performance

Parameter

Linearity range

 Precision

WBC

0.5~99.9x10/L

CV<2.0%

RBC

0.2~9.99x10/L

CV<1.5%

HGB

1.0~299.9 g/L

CV<1.5%

MCV

40~200.0 fL

CV<0.5%

PLT

0~ 999x10/L

CV<4.0%

Carryover

WBC < 0.5 %, RBC < 0.5 %, PLT < 2 %

Operating Environment

Temperature 15 degree C ~ 30 degree C
Humidity 10% ~ 85 %

Prinout

Thremal recorder, printer
Optional external printer

Display

Large color LCD display, resolutions 640 x 480 dot

Power Requirement 175-265v ~ 50 / 60HZ

Weight

25 Kg

     